NORDA Pure Steam Sampler
DSF-220
Steam quality testing solutions !
GMP Compliance
The pure steam coil is made of 316L, with a fully hygienic design, which meets the requirements of GMP for pharmaceutical equipment.
DSF-200 series Pure Steam Sampler is the air-cooled sampler mainly used for the condensation sampling of pure steam. It cools the pure steam in the stainless steel conduit with cold air to condense into water. By inspecting the quality of the sampling water, it is verification whether the pure steam meets the testing standards. DSF-200 Pure Steam Sampler is suitable for pure steam sampling in biological and pharmaceutical pure steam sterilization systems.
The instrument meets the GMP requirements, and the components in contact with pure steam are made of 316L stainless steel, which meets the GMP requirements for pharmaceutical utensils. After sampling, the condensed water is analyzed for microorganisms, conductivity, TOC, and endotoxins.
Pure Steam Sampler is designed for efficient pure steam condensation. stainless steel surface and fully hygienic design meet GMP requirements for pharmaceutical industry!
Pure air-cooled design
Easy Operation
The shell is made of high-quality SUS316 surface brushed material, with a flat and smooth outer material that is not easily contaminated and corrosion-resistant
After fully charged, working time not less than 8h. Battery is easy to disassemble and can be optionally equipped with an additional one
4.3-inch touch screen,display temperature, sterilization & blowing progress
Real-time display of condenser wall temperature
Cleaning function: After sampling, recommended to blow-dry the residual water vapor in the internal pipeline to prevent the growth of bacteria inside it
Portable design
Small in size, practical and portable
Equipped with a pull handle and roller wheels for easy transfer of samples between different points
Pure air cooling design principle,No need to add cooling liquid
Optional special 316 stainless steel steam connection hose
Optional adjustable bracket of sampling outlet for fixing the sampling container
Heat Exchange Area
|
2.9m2
|
Design pressure
|
0.8Mpa
|
Steam pipe
|
8mm*1mm
|
Dimension(cm)
|
15*35*37.5
|
Weight
|
10.5KG
|
Material
|
304 stainless steel (internal direct contact of pure steam is 316 stainless steel )
|
Sampling capability
|
200ml/min
|
Optional
|
Connection hose and others
|
Following requirements for pure steam of the Chinese《Pharmaceutical GMP Inspection Guide》
1,Microbial Limit, the same as inject able water 2,TOC, the same as injectable water
3, Conductivity, the same as inject-able water 4,Endotoxin, 0.25EU/ml (if used for inject able